DOI: 10.19102/icrm.2025.16064
DAVID FRITZ,1,* BEN OSE, MPH,1,* HANNAH ZERR,1 MACI CLARK,1 CAROLINE TRUPP,1,† AMULYA GUPTA,2 AHMED SHAHAB, MBBS, MPH,2,3 SETH H. SHELDON, MD,2 and AMIT NOHERIA, MBBS, SM2
1The University of Kansas School of Medicine, Kansas City, KS, USA
2Department of Cardiovascular Medicine, The University of Kansas Medical Center, Kansas City, KS, USA
3Department of Cardiology, Southern Illinois University School of Medicine, Springfield, IL, USA
![]() |
Download PDF |
|
ABSTRACT. Left bundle branch area pacing (LBBAP) may mitigate pacing-induced cardiomyopathy (PICM) and is increasingly favored over traditional right ventricular pacing (RVP). We sought to evaluate the impact of a practice-wide switch from RVP to LBBAP. We switched practice from RVP to primarily LBBAP at our center in 2020. A retrospective review was conducted to compare patients who underwent LBBAP from 2020–2023 with controls who underwent RVP from 2018–2019. The LBBAP (n = 288; age, 73.3 ± 10.7 years; left ventricular ejection fraction [LVEF], 56.9% ± 11.4%) and RVP (n = 172) groups were similar in terms of age, body mass index, hypertension, diabetes, and LVEF. The LBBAP group as compared to the RVP group had fewer women (38% vs. 51%; P = .006) and longer intrinsic conducted QRS durations (117 ± 28 vs. 110 ± 30 ms; P = .04). LBBAP devices required longer implant (102 vs. 67 min) and fluoroscopy (9.3 vs. 6.9 min) times but resulted in shorter paced QRS durations (122 ± 20 vs. 145 ± 24 ms; all P < .0001). At 3 months, LBBAP patients had higher sensing (13.8 ± 6.1 vs. 12.0 ± 5.6 mV; P = .007), lower pacing impedance (543 ± 98 vs. 576 ± 150 Ω; P = .008), and similar capture threshold (0.78 ± 0.24 vs. 0.76 ± 0.35 V; P = .5) values. Device-related adverse events were similar between the groups (LBBAP 8.7% vs. RVP 8.8%; P = 1.0), which included ventricular lead dislodgement (2.1% vs. 0.6%; P = .3). There were no differences in hazard rates of all-cause mortality (P = .5) or heart failure (HF) hospitalizations (P = .07). In a subgroup of patients with ≥20% ventricular pacing, the average LVEF change during follow-up in the LBBAP group as opposed to the RVP group was +1.6% ± 12.9% versus −3.8% ± 12.0% (P = .03), the average left ventricular internal diameter at end-diastole change was −0.18 ± 0.73 cm versus +0.16 ± 0.45 cm (P = .006), and there were no differences in the hazard rate of all-cause mortality (P = .6) or HF hospitalizations (P = 1.0). Our results suggest there were no adverse consequences of the practice-wide switch from RVP to LBBAP. LBBAP was associated with longer procedure and fluoroscopy times but resulted in narrower paced QRS durations and less PICM.
KEYWORDS. Conduction system pacing, left bundle branch area pacing, pacemaker, pacing-induced cardiomyopathy.
*Both authors contributed equally to this manuscript
†Posthumous.
Dr. Sheldon reports honoraria (modest) from Medtronic and Boston Scientific and speaker/speaker’s bureau compensation (modest) from Medtronic. The remaining authors report no conflicts of interest for the published content. This work was supported by funding from the Reed-Kindred Cardiovascular Fellowship to medical students D. Fritz and B. Ose and from the Kathy Roberts Batenic Summer Fellowship to medical students H. Zerr, M. Clark, and C. Trupp. The abstract of this work was accepted for poster presentation at the 74th Annual Scientific Session of the American College of Cardiology, March 29–31, 2025, in Chicago, IL.
Manuscript received September 13, 2024. Final version accepted February 11, 2025.
Address correspondence to: Amit Noheria, MBBS, SM, 3901 Rainbow Blvd., Kansas City, KS 66160, USA. Email: noheriaa@gmail.com.
A pacemaker is an effective treatment for cardiac conduction disease and is often implanted for irreversible symptomatic bradycardia or pathologic atrioventricular block. Traditionally, ventricular pacing has been accomplished by myocardial capture from a pacing lead placed in the right ventricular (RV) endocardium without targeting the conduction system. However, single-site RV pacing (RVP) can cause dyssynchronous left ventricular (LV) activation, leading to an increased risk for heart failure (HF).1,2 Consequently, a drop in the LV ejection fraction (LVEF) of <50% (pacing-induced cardiomyopathy [PICM]) occurs in 10%–20% of patients receiving chronic RVP.3,4
Different pacing strategies have been proposed to mitigate or treat PICM. These include biventricular pacing (BiVP) and conduction system pacing. BiVP uses an additional lead, usually in an epicardial LV vein, for dual-site pacing across the LV to decrease the ventricular dyssynchrony. However, BiVP implantation is a longer, more complex procedure with challenges and complications that include coronary sinus dissection, high capture thresholds, phrenic nerve stimulation, suboptimal lead location, lead dislodgement, non-physiological electrical activation from the epicardium, and the need for an extra pacing lead.5 Consequently, BiVP is routinely offered only to patients with pre-existing LV dysfunction.6
Conduction system pacing started out as His-bundle pacing (HBP) but is now implemented primarily with left bundle branch area pacing (LBBAP).7 HBP attempts to recruit the bundle branches at the level of the atrioventricular bundle of His and leverage the intrinsic Purkinje network to restore synchronous ventricular activation via a narrow normal QRS complex.8 While HBP may be physiologically superior to RVP and BiVP, the procedure itself is technically difficult. The His bundle is difficult to recruit and may be unsuccessful in 10%–20% of cases.9 Further, the long-term stability of HBP leads can be unpredictable, and concerns remain about the future development of bundle branch blocks below the level of His bundle capture.
LBBAP mitigates such issues by using a lead implanted from the right and penetrated into the interventricular septum to stimulate the left bundle branches. By directly or indirectly recruiting the native LV electrical system, it enables the spread of depolarization synchronously throughout the LV and the QRS width decreases. This is putatively a superior strategy for patients undergoing permanent ventricular pacing.10 LBBAP has been shown to reverse PICM.11 LBBAP, in comparison to BiVP, has demonstrated shorter QRS complex durations, better intraventricular synchrony, and a modest hemodynamic improvement.12,13 Further, as compared to HBP, LBBAP results in lower pacing thresholds, higher R-wave amplitudes, shorter procedure times, and greater implant success.14–16
Despite widespread clinical adoption, there is an absence of comparative randomized clinical trial data on the short- and long-term outcomes of LBBAP as compared to RVP. Not enough is known about the impact of switching routine practice from RVP to LBBAP in a real-world setting. We evaluated the outcomes of a practice-wide shift from conventional RVP to LBBAP at our center.
Our clinical practice of permanent pacemaker implants at a single large tertiary care center, The University of Kansas Medical Center, Kansas City, changed from primarily RVP to primarily LBBAP in 2020. A retrospective chart review was done to assess patients ≥18 years of age who underwent an LBBAP lead implantation between 2020 and 2023 compared to controls who underwent RVP from 2018–2019. The control patients were collected as part of a different retrospective study comparing venous access techniques for cardiac implantable electronic device implantation.17
Baseline patient and procedural characteristics were retrospectively recorded. Implants were classified as either de novo implants or upgrades/revisions (with or without lead extraction). A device-related adverse event was defined as implant-related deep venous thrombosis, infection at the site of implantation, any lead dislodgement, any lead failure, any lead perforation, pocket hematoma requiring evacuation or interruption of anticoagulation, pneumothorax, or pericardial effusion/tamponade requiring pericardiocentesis. Post-procedural follow-up data from multiple electrocardiograms (ECGs), device interrogations, and echocardiograms were recorded. As per standard practice, patients undergoing LBBAP received a 12-lead ECG the morning after device implant to record the paced QRS morphology. We compared the paced QRS duration and LV activation time (LVAT) between LBBAP patients and RVP controls. Data from device interrogations were used to compare LBBAP electrical parameters, such as sensing, pacing impedance, and capture thresholds, to those in RVP controls. Echocardiography was used to determine LVEF, LV internal diameter at end-diastole (LVIDd), LV internal diameter at end-systole, LV end-diastolic volume, and LV end-systolic volume at the time of implant and during follow-up.
Subgroup analyses were performed for patients with ≥20% ventricular pacing burden documented on either 3- or 12-month post-implant device interrogation reports.
Patient data were retrospectively entered and stored securely in an electronic database (Research Electronic Data Capture [REDCap]). Analysis was conducted using JMP Pro version 15 (SAS Institute, Cary, NC, USA). Continuous variables were expressed as mean ± standard deviation values. Categorical variables were expressed as frequency and percentages. Student’s t test and Fisher’s exact test were used to compare continuous and categorical values, respectively. We considered a two-tailed P value of ≤.05 to indicate statistical significance.
Baseline characteristics
There were 288 patients who underwent LBBAP (73.3 ± 10.7 years) between 2020 and 2023 performed by seven operators. These patients were compared to 172 RVP controls (71.8 ± 11.4 years) who underwent implantation between 2018 and 2019, also by seven operators. Baseline patient characteristics can be seen in Table 1. There were no significant differences in age or body mass index, but fewer women underwent LBBAP (38% vs. 51%; P = .006). Patients who underwent LBBAP had longer average intrinsic QRS durations compared to RVP controls (117 ± 28 vs. 110 ± 30 ms; P = .04). There was no difference in baseline medical comorbidities, including hypertension, diabetes, chronic obstructive pulmonary disease, obstructive sleep apnea, and coronary artery disease. There were no statistical differences in medical therapy between the two groups except for the higher use of sodium–glucose cotransporter 2 inhibitors (P = .003) among LBBAP (7.3%) versus RVP (1.2%) patients, as shown in Supplementary Table 1. Patients in both groups had similar baseline LVEFs (56.9% ± 11.4% vs. 56.3% ± 9.8%; P = .7).
|
|
Procedural characteristics
These results are shown in Table 2. Both groups received predominantly de novo device implants instead of upgrades or revisions, though there were fewer de novo implants performed in the LBBAP group compared to the RVP group (91% vs. 98%; P = .0008). The RVP lead location in the control group was classified by the implanting operator as apical in 9 (5.2%) and as non-apical in the remaining 163 (94.8%) patients. The models of the leads used for the LBBAP and RVP groups are provided in Supplementary Table 2. There were similar numbers of implantable cardioverter-defibrillators (7% vs. 3%; P = .06) but more cardiac resynchronization therapy generators (9% vs. 2%; P = .001) implanted in the LBBAP group compared to the RVP group. Among de novo single- or dual-chamber device implantations, procedure durations (102 ± 37 vs. 67 ± 30 min; P < .0001) and fluoroscopy times (9.3 ± 8.7 vs. 6.9 ± 8.1 min; P < .0001) were longer in the LBBAP group than in the RVP group.
|
|
Electrocardiogram result and electrical function
Ventricular paced ECG data are shown in Table 2. Day 1 post-implant ECGs documenting the ventricular paced QRS duration were available for review in 246 (85%) and 76 (44%) patients, respectively, in the two groups. LVAT was able to be measured (presence of the R-wave in lead V6) in 236 (82%) and 34 (18%) patients. When available, the LBBAP leads as compared to the RVP controls were associated with shorter paced QRS durations (121.5 ± 20.0 vs. 144.9 ± 23.7 ms; P < .0001) and LVATs (73.8 ± 20.1 vs. 88.0 ± 23.7 ms; P = .002). Table 3 shows the electrical parameters of the pacing leads in the two groups. On day 1 post-implant, LBBAP leads had similar sensing (12.3 ± 5.7 vs. 11.3 ± 5.7 mV, P = .1) and pacing impedance (625 ± 106 vs. 629 ± 127 Ω, P = .7) values but lower capture thresholds (0.47 ± 0.19 vs. 0.55 ± 0.23 V; P = .0004) compared to RVP leads; however, the difference in capture threshold disappeared at 3-month follow-up (0.78 ± 0.24 vs. 0.76 ± 0.35 V; P = .5). Additionally, at 3 months, the sensing was higher with LBBAP than with RVP (13.8 ± 6.1 vs. 12.0 ± 5.6 mV; P = .007) and the pacing impedance was lower (543 ± 98 vs. 576 ± 150 Ω; P = .008). The results of procedural characteristics and ECG are summarized in Figure 1.
|
Figure 1: (A) Intrinsic QRS duration, (B) paced QRS duration, (C) implant procedure duration, and (D) fluoroscopy time for the entire study population (n = 460). Abbreviations: LBBAP, left bundle branch area pacing; RVP, right ventricular pacing. |
Device-related adverse events
Adverse events related to the devices are summarized in Table 4 and Figure 2. The overall device-related adverse event rate was similar between the LBBAP and RVP (8.7% vs. 8.8%; P = 1.0) groups. The most common adverse event was any lead dislodgement, with 10 (3.5%) dislodgements observed in the LBBAP group and 4 (2.3%) observed in the RVP group. In the LBBAP subset, 6 out of 10 dislodgements involved the LBBAP lead and 4 involved the right atrial lead. In the RVP subset, one out of four involved the RVP lead and three involved the right atrial lead. There were no statistical differences in any specific device-related adverse event or 30-day outcomes between the two groups.
|
Figure 2: Summary of overall device-related adverse events for the entire study population (N = 460). Abbreviations: LBBAP, left bundle branch area pacing; RVP, right ventricular pacing. |
Long-term outcomes
Both groups were followed up to evaluate long-term outcomes (Table 4). The LBBAP group had a shorter follow-up period than the RVP group (17.1 ± 6.4 vs. 59.2 ± 8.9 months; P < .0001). During this shorter follow-up, there were fewer all-cause deaths noted in the LBBAP group, with 25 (8.7%) recorded versus 49 (28.5%) in the RVP group. There was no difference in the number of HF hospitalizations, with 24 (8.3%) versus 19 (11.1%) respectively noted in the LBBAP and RVP groups. Time-to-event Cox proportional hazard ratios of clinical events during follow-up showed no significant difference in the hazard rate of all-cause mortality or HF hospitalizations between the two groups (Table 5).
|
Subgroup analysis of patients with ≥20% ventricular pacing
We evaluated a subgroup of patients with ventricular pacing >20% at either 3-month or 12-month follow-up to evaluate patients at greater risk for PICM (Table 6 and Figure 3). This subgroup included 151 (52%) LBBAP cases and 58 (34%) RVP controls. There were no significant differences in the baseline clinical or echocardiographic characteristics, while baseline ECG measurements revealed a shorter intrinsic conducted QRS duration (121.2 ± 27.2 vs. 143.5 ± 34.1 ms; P = .002) and shorter paced QRS duration (123.8 ± 20.9 vs. 149.7 ± 21.5 ms; P < .0001) in the LBBAP group compared to RVP controls.
|
Figure 3: (A) Intrinsic QRS duration, (B) paced QRS duration, (C) change in LVEF, and (D) change in LVIDd for patients with ≥20% ventricular pacing (n = 209) during follow-up. Abbreviations: LBBAP, left bundle branch area pacing; LVEF, left ventricular ejection fraction; LVIDd, left ventricular internal dimension at end-diastole; RVP, right ventricular pacing. |
The follow-up duration in patients with ≥20% ventricular pacing was shorter for LBBAP (16.0 ± 6.4 vs. 45.1 ± 19.6 months; P < .0001). The change in LVEF (minimum LVEF of 3-, 12-, 24-, and 36-month follow-up referenced to baseline LVEF) in the LBBAP group was +1.6% ± 12.9% compared to that in the RVP group (−3.8% ± 12.0%; P = .03). Similarly, LBBAP was also associated with LVIDd change by −0.18 ± 0.73 cm versus +0.16 ± 0.45 cm, respectively (P = .006). Overall, during the differential follow-up durations, LBBAP patients had lower rates of all-cause mortality (9.6% vs. 46.6%, P < .0001) and HF hospitalization (8.6% vs. 19.0%; P = .05), but there were no statistical differences in time-to-event proportional hazards models.
LBBAP is an attractive alternative to RVP with increased adoption in clinical practice. Ours is one of the first studies describing the experience of a practice-wide shift from RVP to LBBAP.
Key findings
First, LBBAP was associated with lower paced QRS durations, shorter LVATs, lower pacing impedance, and higher sensing compared to RVP. Second, LBBAP was not associated with a decline in LVEF as was the case with RVP. Third, LBBAP implantation resulted in longer total implant duration and fluoroscopy times compared to RVP. Finally, despite higher complexity, LBBAP was not associated with a higher rate of device- or lead-related adverse events.
Baseline differences
There were fewer upgrades/revisions in the RVP controls (2%) compared to the LBBAP group (9%), partly due to the systematic exclusion of patients with pre-existing device leads from the RVP controls who were collected for an unrelated prior research project on ultrasound-guided axillary vein access.17 Although the treated population in both groups was relatively homogenous, there were two key differences. In the LBBAP group, there were fewer women (38% vs. 51%), and the intrinsic conducted QRS duration was also slightly longer (117 ± 28 vs. 110 ± 30 ms). The electrical properties of the heart vary between men and women, potentially influencing the interpretation of our results. For instance, previous studies have indicated that women typically have shorter QRS durations and higher atrial thresholds compared to men.18,19 Additionally, research on leadless pacemakers has demonstrated that women exhibit higher impedance than men.20 However, these differences are generally minor, with various studies presenting conflicting findings, and are likely to have had only a minimal impact on the observed outcomes.
Procedural characteristics
Several procedural variables differed between the two groups. In the LBBAP group, a greater proportion of implants were upgrades/revisions rather than new de novo implants compared to those in the RVP group (9% vs. 2% new implants). Previous studies have shown no difference in clinical response between upgrades and de novo implantation, suggesting that the uneven distribution of implantation types has minimal impact on the interpretation of our study.21
As anticipated, the paced QRS duration was much shorter in the LBBAP group compared to RVP controls, indicating the recruitment of the native conduction system. Further, average fluoroscopy times and implant durations were longer in LBBAP patients compared to RVP controls, which is replicated in previous studies and is reflective of the more complex nature of LBBAP implantation.22 Extended fluoroscopy times during procedures introduce risks associated with higher radiation doses for both operators and patients. Operators can potentially mitigate this risk by reducing the fluoroscopy frame rate, using ultrasound-guided axillary venous access, or employing techniques such as echocardiography or electroanatomic mapping when suitable.23,24
Device-related adverse events and outcomes
No statistically significant differences were found in device-related adverse events, 30-day outcomes, 1-year mortality, and HF hospitalizations between the LBBAP and RVP cohorts. In contrast, a 2023 meta-analysis comparing LBBAP with RVP revealed that, while total lead-related complications were similar in both groups, the LBBAP group exhibited lower all-cause mortality and a reduced risk of HF hospitalization over a 16-month follow-up period.25 In our analysis, the hazard rates of HF hospitalization or death were not different in the two groups.
Electrical parameters
Compared to RVP, the capture threshold was slightly lower in the LBBAP group on day 1 post-implant, although the difference vanished at 3 months post-implant. A meta-analysis pooling data from 25 studies revealed no discernible difference in the capture threshold between the two groups immediately post-procedure, suggesting that the small disparity observed in our data may be a random artifact rather than a substantive trend.25 Further, consistent with the findings from previous studies, at 3 months post-implant, the LBBAP group had higher sensing and lower impedance than the RVP group, which can be attributed to the placement of leads in the muscular interventricular septum rich in conduction fibers.25,26
Patients with ventricular pacing ≥20%
We examined patients with higher ventricular pacing burdens (≥20%) separately to evaluate its effect on cardiac function. Patients in the LBBAP group were more likely to receive ≥20% ventricular pacing (52% vs. 34%). This may be reflective of the operators’ avoidance of RVP (preference for BiVP) in patients likely to receive a high burden of ventricular pacing and stringent programming to avoid ventricular pacing in those receiving RVP as opposed to LBBAP. Both the groups were similar in baseline characteristics and echocardiographic variables, although the intrinsic conducted QRS was shorter in the LBBAP group as compared to the RVP group. This baseline difference is again likely due to the perceived safety with LBBAP, which permits operators to pace more liberally even in cases without atrioventricular blocks. Additionally, among the ≥20% ventricular paced population, the median ventricular pacing burden at 3 months was greater in the LBBAP group compared to the RVP group (98.8% vs. 90.0%). In line with previous studies, the paced QRS duration was shorter in LBBAP than in RVP.25
Among the subgroup with ≥20% ventricular pacing, the LBBAP patients had a favorable echocardiographic response, with the average LVEF improving and the average LVIDd decreasing as compared to vice versa in the RVP group. However, it is worth noting that the RVP controls for this subgroup analysis had a longer baseline QRS duration and thus may have been predisposed to developing cardiomyopathy. These findings can be interpreted to suggest that a wider baseline QRS duration may portend a greater chance of systolic cardiomyopathy and potential need for cardiac resynchronization therapy, and therefore an even stronger reason to consider LBBAP over RVP.
Overall follow-up with ventricular pacing ≥20%
During the entire duration of follow-up, the LBBAP group had lower all-cause mortality, fewer HF hospitalizations, and fewer days hospitalized for HF; however, this difference is attributed to the large difference in the duration of follow-up between the two groups. Indeed, there were no statistical differences in the hazard rates for mortality or HF in the proportional hazards models that account for the differences in follow-up duration.
Limitations
Our analysis is subject to several limitations. First, all typical limitations of a retrospective study apply, and our results may be biased by systematically missing data, especially missing outcomes of ECG and electrical parameters in the RVP group. Second, the procedures were conducted by seven operators without standardized protocols or endpoints. Third, the use of a historical cohort with different durations of follow-up affected the analysis of long-term outcomes. Finally, the imbalances in sex and device distribution between the two groups could have affected outcomes and electrical parameter data.
LBBAP is increasingly recognized as a viable alternative to RVP. Key findings include a lower paced QRS duration, lower pacing impedance, and higher sensing in the LBBAP group, despite the procedure’s increased complexity and longer implant times. Importantly, LBBAP did not result in any higher rates of device-related adverse events, including lead dislodgements. Our analysis confirms the benefits of LBBAP, and we did not observe any adverse consequences from a practice-wide shift from RVP to LBBAP.