Use of TYRX Antibacterial Envelope Associated with Extremely Low 90-Day Cardiac Device Infection Rates.
New Prospective Data Presented on 1,000 Patients at 2013 Heart Rhythm Society Shows >90% Fewer Major Device Infections
Dr. Charles Henrikson, the Chief of Electrophysiology at the Oregon Health Sciences University, presented new prospective data on 1,000 patients showing 95% fewer major CIED infections in patients undergoing CIED replacement procedures using the AIGISRx Antibacterial Envelope compared to similar high-risk cohorts. These results were presented during the Late Breaking Clinical Trials session at Heart Rhythm 2013.
CITADEL / CENTURION is a prospective, multicenter clinical study to evaluate the major device infection and mechanical complication rates in the 12 months after implantation, in patients at high risk for CIED infection who have their CIED implanted with an AIGISRx Antibacterial Envelope. Study patients were enrolled at 55 US centers, and were at high-risk for infection because they were undergoing a CIED replacement procedure with an implantable cardioverter-defibrillator (ICD), (CITADEL), or a cardiac resynchronization therapy (CRT) device (CENTURION).
Key study findings include:
- The CITADEL / CENTURION cohort at 90 days of follow-up had 95% fewer major CIED infections than the pre-defined published control cohort of 533 ICD and CRT replacement procedures (Gould et al. JAMA 2006, 295(16); 1907-1911) which had a major CIED infection rate of 1.88% at a mean follow-up of 81 days (0.1% vs. 1.88%; P<0.001).
- The CITADEL / CENTURION cohort at 90 days of follow-up had 94% fewer major infections than the 45-day major infection rate of 1.7% reported for the cohort of 1081 ICD/CRT replacement procedures in the Ontario ICD Database (Krahn et al. Circulation Arrhythmia and Electrophysiology 2011 4(2) 136-42 (0.1% vs. 1.7%; P<0.001).
- There was 1 major infection (0.1%), the primary efficacy endpoint of the study, after 90 days of follow-up. There were 11 minor infections (limited to the incision and skin) (1.1%).
- The incidence of the most common mechanical complication, major hematomas (1.5%), was not significantly different than the pre-defined control cohort (2.3%; P = NS).
- There were no unanticipated serious AIGISRx-related adverse events. There were 20 (2%) deaths, none related to the AIGISRx.