Failure of Cardiac Resynchronization Therapy due to Inappropriate Inhibition of Left Ventricular Pacing during Biventricular Pacing

DOI: 10.19102/icrm.2015.060305

ROBERT H. HELM, MD, ADVAY G. BHATT, MD and KEVIN M. MONAHAN, MD

Boston University School of Medicine, Boston, MA

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.Cardiac resynchronization therapy (CRT) has been shown to be an effective therapy in patients with cardiomyopathy and left bundle branch block; however, 20% to 30% of patients receiving CRT fail to respond, and this maybe in part due to poor therapy implementation. Recent data suggest that a basal left ventricular (LV) lead location is more efficacious than apical placement. We describe a case in which CRT exacerbated heart failure because of the combined effect of a basal lead location and left ventricular sensing algorithm. Disabling this sensing algorithm improved CRT delivery.

KEYWORDS.biventricular pacing, cardiac resynchronization therapy, left atrial far-field oversensing, left ventricular lead placement, left ventricular protection period.

The authors report no conflicts of interest for the published content.
Manuscript received December 11, 2014, final version accepted January 14, 2015.

Address correspondence to: Robert Helm, MD: 88 East Newton Street/Collamore 8, Boston, MA 02118. E-mail: Robert.helm@bmc.org

Introduction

Cardiac resynchronization therapy (CRT) has emerged as a highly effective therapy in patients with cardiomyopathy and left bundle branch block.13 Unfortunately, 20% to 30% of patients receiving the therapy fail to demonstrate clinical improvement.4,5 Basal left ventricular (LV) lead location has been shown to be more efficacious than apical.6 We describe a case where implementation of CRT was detrimental because of the combined effect of basal lead location and left ventricular sensing.

Case report

A 65 year-old male with non-ischemic cardiomyopathy was referred for upgrade of his implantable cardioverter-defibrillator (ICD) to a biventricular system for progressive heart failure despite an optimized heart failure regimen. He was quite functionally limited and had New York Heart Association class III symptoms. His electrocardiogram showed non-specific intraventricular conduction delay with a QRS duration of 166 ms.

He was electively brought to the electrophysiology laboratory, and the pocket was opened. The coronary sinus was accessed, and a balloon-occlusive venography of the coronary sinus showed a small lateral branch of the coronary sinus. A Medtronic 4136-88 (Medtronic Inc, St. Paul, MN) was advanced into the lateral branch vessel, and suitable lead parameters were found. The pacing impedance was 1,146 ohms, and the capture threshold was 0.4 volts at 0.5 ms. There was no phrenic nerve or atrial capture with high-output pacing. The lead connected to a Boston Scientific COGNIS TM 100-D generator (model # N119; Boston Scientific Corp., Boston, MA), and the wound was closed. The following day, posterior-anterior and lateral chest renographs showed stable basal LV lead position (Figure 1a,b). Interrogation of the device also demonstrated stable LV lead parameters, and the patient was discharged.

Two weeks later, he was admitted to our institution with acute decompensated heart failure. A chest renograph showed stable LV lead position (Figure 1c,d). Interrogation of his ICD showed unchanged LV lead parameters. The ventricular pacing burdens were 66% and 82% for the left ventricular (LV) and right ventricular (RV), respectively. The discrepancy was due to inhibition of LV pacing during DDD pacing (Figure 2a). When pacing was inhibited, two electrical potentials were observed on the LV channel that corresponded to atrial and ventricular activation (Figure 2b). The device sensed the first component (far-field left atrial component) as LV activation and inhibited the scheduled LV pace event to avoid pacing in the vulnerable recovery period. The scheduled RV pace event was delivered. We concluded that the lack of ineffective biventricular pacing and burden of RV-only pacing contributed to heart failure exacerbation in our patient. Disabling the LV sense feature restored effective biventricular pacing.

crm-06-03-1954-f1.jpg

Figure 1: Posterior-anterior and lateral chest renographs on the day following LV lead placement (a, b) and 2 weeks later (c, d). The black and white arrows mark the proximal and distal poles of the LV lead, respectively. The LV lead position is stable between the initial and follow-up exams and is clearly beyond the body of the coronary sinus.

crm-06-03-1954-f2.jpg

Figure 2: a. Intracardiac electrograms recorded from the atrial (A), right ventricular (RV), and left ventricular leads (LV) during DDD pacing. A marker channel indicates atrial sensing (AS), inhibition of LV pacing (L'Inh-LVP) and right ventricular pacing (RVP). The LV electrogram showed two components (left atrial far-field and local LV). The left atrial far-field electrogram is sensed and corresponds to inhibition of the LV pace event (L'Inh-LVP on marker channel). b. Intracardiac electrograms recorded when pacing is inhibited (VVI pacing mode). The LV electrogram showed far-field left atrial activation (black arrow) and LV activation (gray arrow). The far-field left atrial activation corresponds to an LV sensed event (LVS on marker channel). LV activation is not sensed due to a 300-ms blanking period following an LV sensed event (no corresponding marker).

Discussion

Identifying causes of non-response to CRT can be challenging. Assuring adequate therapy implementation is important before classifying the patient as a non-responder. The present case report highlights an unusual situation where device programming and lead placement led to poor CRT implementation.

Left atrial far-field sensing has been previously reported in CRT systems that shared a common ventricular port (i.e., RV and LV leads connected via a Y-adaptor).7,8 In conventional CRT systems with dedicated LV ports, left atrial sensing has only been reported in a patient with new onset atrial flutter that resolved with restoration of sinus rhythm.9 Our case is the first to demonstrate this phenomenon occurring in sinus rhythm, and this situation may not be as easily recognized as atrial flutter.

Pacing within a chamber without sensing has the theoretical risk of arrhythmia induction. Boston Scientific cardiac resynchronization devices have independently programmable RV and LV pacing and sensing. While RV sensing is used for RV-based timing and RV pacing inhibition, LV sensing is only used to inhibit a scheduled LV pace event. There have been rare case reports of CRT having a pro-arrhythmic effect.10,11 Thus, following a sensed LV event, Boston Scientific incorporated a programmable protection feature known as the Left Ventricular Protection Period (LVPP). The LVPP inhibits a scheduled pace event for 300 to 500 ms to avoid LV pacing during the vulnerable recovery period (T-wave). This feature is independent of RV sensing and does not reset the lower rate timer. Among other cardiac device manufacturers including Medtronic, St. Jude, and Biotronik, Biotronik is the only corporation that has a similar LV sense feature, and it is referred to as LV T-wave protection.

Basal LV lead location is desirable, and in our patient it resulted in a dual-component electrogram corresponding to atrial and ventricular activation. With the LVPP parameter turned “ON,” the left atrial far-field electrogram was sensed as an LV event, and the scheduled LV pace event was inhibited during biventricular pacing. The RV pace event was not inhibited. This resulted in resynchronization failure, a high burden of RV pacing, and acute decompensated heart failure. Disabling the LVPP feature, which is nominally turned “ON,” achieved appropriate biventricular pacing with simultaneous LV and RV pace events. An alternative treatment strategy would have been to change the LV sense configuration to an extended bipolar configuration using the distal LV dipole and RV coil. The distal dipole is further from the mitral annulus and is less likely to record atrial activity (Figure 3). Inhibition of LV pacing during CRT pacing was not observed with the extending LV bipolar configuration in sinus rhythm.

crm-06-03-1954-f3.jpg

Figure 3: A comparison of bipolar and extended bipolar LV sensing configurations. Unlike true bipolar sensing (LV tip to ring), far-field left atrial sensing was not observed with extending LV bipolar sensing (LV tip to RV coil). Inhibition of LV pacing during CRT pacing was not observed with the extending LV bipolar configuration in sinus rhythm.

Conclusion

In conclusion, a basal LV lead location with far-field left atrial sensing in the presence of LV sensing (LVPP) may lead to inappropriate inhibition of LV pacing and poor implementation of CRT therapy.

References

  1. Bristow MR, Saxon LA, Boehmer J, et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 2004; 350:2140–2150. [CrossRef] [PubMed]
  2. Abraham WT, Fisher WG, Smith AL, et al. Cardiac resynchronization in chronic heart failure. N Engl J Med 2002; 346:1845–1853. [CrossRef] [PubMed]
  3. Cleland JG, Daubert JC, Erdmann E, et al. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 2005; 352:1539–1549. [CrossRef] [PubMed]
  4. Birnie DH, Tang AS. The problem of non-response to cardiac resynchronization therapy. Curr Opin Cardiol 2006; 21:20–26. [CrossRef] [PubMed]
  5. Yu CM, Wing-Hong FJ, Zhang Q, Sanderson JE. Understanding nonresponders of cardiac resynchronization therapy–current and future perspectives. J Cardiovasc Electrophysiol 2005; 16:1117–1124. [CrossRef] [PubMed]
  6. Singh JP, Klein HU, Huang DT, et al. Left ventricular lead position and clinical outcome in the multicenter automatic defibrillator implantation trial–cardiac resynchronization therapy (MADIT-CRT) trial. Circulation 2011; 123:1159–1166. [CrossRef] [PubMed]
  7. Taieb JM, Barnay C, Linde C, Mortensen P, Menardis M. Left atrial far-field sensing by left ventricular leads: A potential hazard in cardiac resynchronization therapy. Europace 2005; 7:7611–7616. [CrossRef] [PubMed]
  8. Oguz E, Akyot A, Okmen E. Inhibition of biventricular pacing by far-field left atrial activity sensing: case report. Pacing Clin Electrophysiol 2002; 25:1517–1519. [CrossRef] [PubMed]
  9. Duparc A, Mondoly P, Detis N, et al. Atrial oversensing by an LV lead during typical flutter: the interest of electronic repositioning. Pacing Clin Electrophysiol 2013; 36:e53–e55. [CrossRef] [PubMed]
  10. Guerra JM, Wu J, Miller JM, Groh WJ. Increase in ventricular tachycardia frequency after biventricular implantable cardioverter defibrillator upgrade. J Cardiovasc Electrophysiol 2003; 14:1245–1247. [CrossRef] [PubMed]
  11. Mykytsey A, Maheshwari P, Dhar G, et al. Ventricular tachycardia induced by biventricular pacing in patient with severe ischemic cardiomyopathy. J Cardiovasc Electrophysiol 2005; 16:655–658. [CrossRef] [PubMed]