Reduction in Cryothermal Energy Delivery During Isolation of the Left Persistent Superior Vena Cava: Implications for Safety and Efficacy

DOI: 10.19102/icrm.2013.040106

JOHN HARDING, MD, CHANTIP WIKLER, RCIS and ROBERT SANGRIGOLI, MD

Division of Internal Medicine/Cardiology, Doylestown Hospital, Doylestown, PA

PDF Download PDF
tweeter Follow Us >>

ABSTRACT. Aims: The left superior vena cava (LPSVC) can be a trigger for atrial fibrillation but can be difficult to isolate, which can increase procedure-related risk. Our aim was to demonstrate whether a reduction in cryothermal energy delivery was feasible while still achieving isolation of the LPSVC, potentially improving safety of ablation in this structure. Methods: We present two cases where isolation of the LPSVC was achieved utilizing cryoballoon technology and an inner recording catheter. Preoperative computed tomography imaging was used to appropriately size the balloon to match the diameter of the LPSVC. We utilized an inner recording catheter to provide immediate feedback on electrical isolation of this structure, with the goal of reducing cryothermal energy delivery. Results: Electrical isolation of the LPSVC was achieved after just over 2 min of cryoablation in Case 1 and in 6 min in Case 2, a substantial reduction compared with prior reports. Conclusion: Minimizing energy delivery in the LPSVC with patient-specific balloon sizing and real-time feedback on electrical isolation is feasible and should protect adjacent vital structures. Such techniques may improve safety and durability of ablation in this area.

KEYWORDS. ablation, atrial fibrillation, cryoablation, left persistent superior vena cava.

Drs. Harding and Sangrigoli receive consulting fees (modest) and research support (modest) from Medtronic. Dr. Harding also receives consulting fees (modest) and research support (modest) from Biosense Webster.
Manuscript received October 21, 2012, final version accepted November 12, 2012.

Address correspondence to: John Harding, MD, Doylestown Hospital, 599 W State Street Doylestown, PA 18901. E-mail: jharding246@comcast.net

Introduction

The left persistent superior vena cava (LPSVC) is an uncommon finding in adults that has been shown by other groups to play a role in atrial fibrillation (AF) initiation.1 The ligament of Marshall, an embryologic remnant of the LPSVC, has also been shown to play a role in AF initiation.2 Electrical isolation of the LPSVC has been reported by other groups, with both radiofrequency energy and cryothermal energy.35 In those reports, however, confirmation of isolation was determined after prolonged cryothermal energy applications, increasing the risk for damage to adjacent structures such as the circumflex artery and left phrenic nerve.5 Given the close proximity of these vital structures to the LPSVC, precise delivery of ablation energy is paramount. Conflicting reports of the safety of ablation in the LPSVC also exist, and the actual complication rate for ablation within the LPSVC is unknown.5

A recent advance in cryoablation technology includes the availability of a circular mapping catheter (AchieveTM; Medtronic Inc., Minneapolis, MN) that can be used in conjunction with the cryoablation balloon (Arctic FrontTM, Medtronic Inc.). The AchieveTM allows real-time recording of isolation of potentials at a more distal location to the cryothermal ablation site. The time to isolation of pulmonary vein (PV) potential after initiation of cryothermal energy, or “time to effect,” has been shown to significantly correlate with durable isolation of the pulmonary veins and acute procedural success by other groups.6,7

We present two cases demonstrating LPSVC isolation in patients with paroxysmal AF utilizing cryothermal energy via cryoballoon technology. Each patient had continued atrial ectopy or non-sustained atrial tachycardia from the LPSVC/distal coronary sinus after pulmonary vein isolation. Each case illustrates a reduction in overall energy delivery compared with prior reports, and may have implications for safety and durability of electrically isolating the LPSVC.

Methods/results

Case 1

A 59-year-old male with a history of paroxysmal AF and LPSVC (Figure 1a) underwent initial pulmonary vein isolation and limited focal ablation within the LPSVC in 2009 without isolation of that structure. The patient had been free of AF for over 2 years until a late recurrence of AF which subsequent drug therapy failed to control.

crm-04-01-1086-f1.jpg

Figure 1: (a) Preprocedure computed tomography scan for Case 1 noting left superior vena cava (LPSVC) with dilated coronary sinus and connection to right-sided SVC (white oval delineates approximate site of cryothermal lesion). (b) Occlusion of LPSVC in Case 1 with 28-mm cryoballoon. (c) Occlusion of LSPV with 23-mm cryoballoon in Case 2. (d) Occlusion of LPSVC in Case 2; note pacemaker leads coursing through LPSVC/coronary sinus.

At repeat procedure, the left superior and right superior pulmonary veins had segmental reconnection and these were reisolated with cryothermal energy. The left and right inferior veins were chronically isolated and were not ablated. We utilized a “freeze–thaw–freeze” strategy to isolate the pulmonary veins, demonstrating occlusion with pulmonary vein venogram and color Doppler flow with intracardiac echocardiography. Despite durable isolation of the pulmonary veins, non-sustained atrial tachycardia from the LPSVC continued. Cryothermal energy application with a 28-mm cryoballoon was utilized along the anterior left superior pulmonary vein and posterior portion of the left atrial appendage (LAA), causing delay in local LPSVC potential (Figure 2) without electrical isolation of the LPSVC. The 28-mm cryoballoon was chosen based upon the measured diameter of the LPSVC (24 mm) at the carina of the left pulmonary veins and the ostial sizes of the PVs from preprocedure computed tomography (CT) scan. The coronary sinus, LPSVC, and LAA were paced for phrenic nerve capture prior to LPSVC ablation, and phrenic nerve capture could not be elicited. We then placed the 28-mm cryoballoon and AchieveTM catheter into the coronary sinus and performed ablation at the mid-LPSVC (Figure 1b). The first application demonstrated rapid isolation of the LPSVC within 6 s, achieving rapid cooling to –65°C, and energy application was halted after 13 s. The LPSVC reconnected and a second more proximal energy application achieved isolation in 4 s, achieving a temperature of –62°C for lesion duration of 120 s (Figure 2). This led to durable isolation of the LPSVC, and no further ectopic beats were seen from this structure.

crm-04-01-1086-f2.jpg

Figure 2: (a) Initial LSPV and left superior vena cava (LPSVC) recording. (b) After LSPV isolation, LPSVC potential persists. (c) Cryoablation on posterior side of left atrial appendage (LAA) induces delay in LPSVC potential (blue arrow). (d) Isolation of delayed LPSVC potential with cryoablation at mid-LPSVC (blue arrows) and continued far-field LAA recording.

We monitored ST segments throughout freezing applications, and fluoroscopy revealed normal diaphragm excursion post ablation. The patient experienced mild pericarditis symptoms without electrocardiogram findings which responded to oral non-steroidal anti-inflammatory drug therapy. At short-term follow-up (6 months), the patient was AF free.

Case 2

A 75-year-old male with a history of sick sinus syndrome and prior dual-chamber pacemaker placement (Accent RFTM, St. Jude Medical, St Paul, MN; leads, 2088 TCTM, St. Jude Medical) via the LPSVC/coronary sinus presented with highly symptomatic and drug-refractory paroxysmal atrial fibrillation. He was referred for catheter ablation.

During the procedure, PV isolation was achieved with a 23-mm cryoballoon using the technique described in Case 1. Preprocedure CT scan revealed a LPSVC diameter of 21 mm at the level of the carina of the left pulmonary veins and overall smaller pulmonary veins (all <20 mm in diameter). Accordingly, we chose to use the smaller size of balloon. After pulmonary vein isolation, infusion of isoproterenol 3 µg/min induced frequent atrial premature depolarizations with earliest activation at the LPSVC/coronary sinus junction. We again paced LAA and LPSVC, and could not elicit left phrenic nerve capture. Two separate 3-min freezing applications utilizing 23-mm cryoballoon were placed at the LPSVC/coronary sinus junction (Figure 1d), achieving electrical isolation of the mid/distal LPSVC (Figure 3). Temperatures during each of the freezes ranged from –60 to –70°C. Pre- and post-ablation electrograms from the LPSVC demonstrate significant signal reduction and exit block (Figure 3). Pacemaker lead integrity during cryothermal energy application was assessed by pacing just above threshold during ablation. Lead impedance was checked before, during, and after freezing applications (right atrium 440 ohm, range 421–450 ohms; right ventricle 600 ohm, range 600–624 ohms) with stable lead parameters. Normal diaphragm excursion on fluoroscopy was seen post procedure. The patient noted no adverse symptoms post ablation. Device interrogation 6 months post procedure revealed no further AF and stable lead parameters.

crm-04-01-1086-f3.jpg

Figure 3: (a) Signal in LPSVC pre-ablation for Case 2. (b) Pacing at mid-left superior vena cava (LPSVC) demonstrating exit block to the atria with far-field left atrial appendage recording. (c) Electroanatomic voltage map (St. Jude Medical) demonstrating level of entrance block in the LPSVC.

Discussion

To our knowledge, we have demonstrated the first reported electrical isolation of the LPSVC with cryoballoon technology utilizing “real-time” recording. In addition, we have also reported the first case utilizing cryothermal energy in direct apposition to pacemaker leads. The overall reduction in cryothermal energy delivery is likely due to appropriate balloon sizing and feedback from “real-time” recordings from the inner lasso catheter.

Isolation of the LPSVC with cryothermal ablation

Other groups have demonstrated that isolation of LPSVC is feasible but often requires multiple procedures, thereby increasing potential procedural risks.3,5 Furthermore, prior approaches have utilized prolonged cryothermal applications to isolate the LPSVC. In Schneider et al,3 the authors used three distinct 6-min balloon applications, while Wissner et al5 targeted a temperature of –80°C with 5-min balloon applications. Total time of cryothermal energy application was not reported in that study however. Given recent case reports of atrioesophageal fistula formation with the cryoballoon,8 prolonged freezing applications and perhaps lower balloon temperatures are not ideal and may lead to higher risk for damage to adjacent structures.

It is possible that tailoring the size of the cryoballoon based upon preprocedure CT imaging allowed more contiguous lesion formation in our two cases, reducing time of ablation and potentially reducing risk. In Case 1, we demonstrated almost immediate isolation of the LPSVC, although delay in the LPSVC potential through ablation at adjacent structures in the left atrium was already present. We targeted a temperature of –55 to –70°C, and stopped energy delivery in the first application due to rapid cooling to –65°C within 3 s. The second freezing application was applied in the mid-LPSVC with almost immediate isolation of LPSVC. Freezing duration was continued for 120 s with a more gradual cooling profile, reaching a minimum temperature of –62°C. In Case 2, we were able to achieve similar cooling parameters with isolation of the mid/distal LPSVC after two distinct 3-min applications. Real-time recording of isolation in conjunction with appropriate balloon sizing allowed us to minimize energy delivery, which potentially has implications for safety. Taken together, this suggests that cryothermal energy can be limited to electrically isolate the LPSVC, potentially minimizing risk to adjacent structures while achieving procedure endpoints needed for long-term success.

Cryothermal energy application immediate to pacemaker leads

We are unaware of data on cold tolerances of pacemaker leads within the human body, and the long-term consequences of cryothermal application in direct apposition to pacemaker leads. Anticipating the need for LPSVC isolation in Case 2, review of device literature in advance of the procedure provided little guidance regarding the safety of ablation near pacemaker leads. We chose to perform ablation targeting isolation of the distal coronary sinus and LPSVC given isoproterenol provoked ectopic beats from this structure despite durable pulmonary vein isolation. During cryothermal applications, we elected to monitor lead integrity with periodic impedance testing during energy application while pacing in DDD mode at near-threshold output. The lead parameters remained stable throughout each freezing application and post-procedure lead parameters remained unchanged in short-term follow-up (6 months). Case 2 presents a unique challenge given the absence of data on this specific situation, but appears to be feasible and safe in the short term. In the absence of long-term data on cold tolerances of pacemaker leads in the human body, however, cryothermal application in direct apposition to pacing leads cannot be routinely recommended and should be avoided in pacemaker-dependent patients.

Conclusion

Our experience utilizing cryoballoon technology suggests that the LPSVC isolation is feasible and can be isolated rapidly. As a site of AF initiation due to its embryologic relationship to the ligament of Marshall, techniques to improve LPSVC isolation are integral to long-term AF control. Utilizing real-time electrical recordings provides the operator direct feedback on time of isolation reducing duration of energy delivery. Preoperative CT or magnetic resonance imaging can lead to more appropriate balloon sizing, perhaps achieving a more rapid and durable isolation of this structure. Minimizing energy delivery in the LPSVC with patient-specific balloon sizing and real-time feedback on electrical isolation should protect adjacent vital structures and can improve safety of ablation in this area. Further experience is required to build upon the limited reports in the literature utilizing cryothermal energy to achieve isolation within the LPSVC.

Limitations

Given the rarity of encountering LPSVC during AF procedures, comparison of efficacy and safety between radiofrequency energy and cryothermal energy for LPSVC will be difficult. It is also possible that patient-specific balloon sizing, leading to improved contact and lesion formation, may paradoxically increase risk to adjacent structures despite shorter lesion duration. More compliant or alternative balloon and sheath technology perhaps will allow improved lesions of shorter duration, further improving safety of this strategy. We also acknowledge that variability in the size of the LPSVC may limit the usefulness of patient-specific balloon sizing in an extremely dilated or diminutive LPSVC.

It is difficult to extrapolate our findings from two case reports. However, the published literature includes a paucity of patients with a LPSVC that have undergone AF ablation. Our follow-up of 6 months is short and longer term follow-up on both lead integrity in Case 2 and clinical arrhythmia recurrences are needed to further support this strategy.

Acknowledgment

The authors would like to express their appreciation to Mathew Hutchinson, MD for critically reviewing the manuscript.

References

  1. Hsu LF, Jais P, Keane D, et al. Atrial fibrillation originating from the persistent left superior vena cava. Circulation 2004; 109:828–832. [CrossRef] [PubMed]
  2. Hwang C, Wu TJ, Doshi R, Peter CT, Chen P-S. Vein of Marshall cannulation for the analysis of electrical activity in patients with focal atrial fibrillation. Circulation 2000; 101:1503–1505. [CrossRef] [PubMed]
  3. Schneider M, Schade A, Koller ML, Schumacher B. Cryoballoon isolation of paroxysmal atrial fibrillation within the dilated coronary sinus in a case of persistent superior vena cava. Europace 2009; 11:1387-1389. [CrossRef] [PubMed]
  4. Liu H, Lim K-T. Murray C, Weerasooriya R. Electrogram-guided isolation of the left superior vena cava for treatment of atrial fibrillation. Europace 2007; 9:775–780. [CrossRef] [PubMed]
  5. Wissner E, Tilz R, Konstantinidou M, et al. Catheter ablation of atrial fibrillation in patients with persistent left superior vena cava is associated with major intraprocedural complications. Heart Rhythm 2010; 7:1755–1760. [CrossRef] [PubMed]
  6. Chierchia GB, de Asmundis C, Namdar M, et al. Pulmonary vein isolation during cryoballoon ablation utilizing the novel Achieve inner lumen mapping catheter, a feasibility study. Europace 2012; 14:962-967. [CrossRef] [PubMed]
  7. Chun J, Furnkranz A, Metzner A, et al. Cryoballoon pulmonary vein isolation with real time recordings from the pulmonary veins. J Cardiovasc Electrophysiol 2009; 20:1203–1210. [CrossRef] [PubMed]
  8. Stockigt F, Schrickel J, Andrie R, et al. Atrioesophageal fistula after cryoballoon pulmonary vein isolation. J Cardiovasc Electrophysiol. 2012; 23:1254-7. [CrossRef] [PubMed]